Requires at least 500 mm behind the unit and 100 mm on the left side for ventilation.
Comprehensive Guide to the Shimadzu GC-2010 Service Manual: Maintenance, Troubleshooting, and Calibration
If the TCD baseline cannot be zeroed, the filaments may be unevenly aged or damaged. The service manual outlines the process for measuring filament resistance across the Wheatstone bridge circuit to determine if a full cell replacement is required. 5. Electrical Troubleshooting and Board Diagnostics shimadzu gc-2010 service manual
Third-party and "free download" websites may host unofficial or outdated documents. Always verify the part number of any manual you acquire with Shimadzu.
If the flame fails to ignite or the baseline is noisy, remove the FID collector assembly. Use the specialized jet removal tool to unscrew the jet. Inspect it under magnification for cracks or blockages. Clean it ultrasonically in hexane or replace it. Requires at least 500 mm behind the unit
Here are some of the most common maintenance tasks you can perform using the official documentation available for the GC-2010.
While a full manual is irreplaceable, here are a few common GC-2010 issues and their typical solutions: If the flame fails to ignite or the
Shut off all gas flows and allow the internal sensors to stabilize. Register the ambient zero-pressure value.